產品介紹

直流電源供應器
  • Facebook
  • LINE
  • Twitter
  • LinkedIn
DLM 600 Watt - AMETEK Sorensen  Near-linear Programmable DC Power Supply | 洛克Lockinc
功能說明

Sorensen DLM 600 Watt
Near-linear Programmable DC Power Supply that provides continuously variable output voltage and current for use in a compact 1U half-rack chassis

Product Overview

The Sorensen DLM 600 series of programmable power supplies is designed to provide continuously variable output voltage and current for a broad range of applications in a compact 1U (1.75 inches) high, half-rack (8.5 inch) wide chassis.
With the use of Zero Voltage Switching (ZVS) technology, these supplies are able to achieve exceptionally low ripple and noise rivaling larger and more expensive linear power supplies. In addition, the high efficiency and fast load transient response assure that the DLM 600 is ideal for even the most demanding applications.
The DLM 600 models are ideal for high density multiple output rackmount requirements or low profile benchtop applications. Output voltages from 0-5VDC to 0-300VDC and currents from 0-2A to 0-75A are available.
The cooling air intake is at the front and sides with exhaust at the rear and sides. Variable speed fans are controlled as a function of ambient air temperature and load, thus reducing acoustic noise and increasing fan life. This allows supplies to be stacked one directly on top of another without space in between, yielding maximum rackmount packing density and a wide operating temperature of 0-50° C. An optional rackmount kit is available to easily mount one or two units side-by-side.

The front panel layout makes the DLM 600 series easy to use. Voltage and current can be set from individual 10-turn potentiometer knobs. Control push buttons include power on, output on, local/remote, voltage/current preview and overvoltage protection preview. Set-point or actual voltage/ current values can be viewed on two 3 ½digit LED displays.
Supplies can be connected in series or parallel. All models automatically accept any standard single phase input without manual set up.

Features And Benefits

Key Features

600 watts in 1U (1.75 inches) high, half rack (8.5 inches) wide; no top or bottom clearance spacing required.

Near Linear Ripple & Noise

Near Linear Ripple & Noise: Ripple as low as 2.5mV rms, noise as low as 15mV p-p.

Remote Control

  • 16 bit programming through LXI -compliant Ethernet LAN / RS-232C (Option M130)
  • 12-bit programming through IEEE-488.2 / RS-232C (Option M9G)
  • Standard analog programming 0-5V, 0-10V, or 0-5kΩ.
  • isolated analog programming (option)

Pushbutton Preview

Overvoltage protection, voltage and current preview buttons allow viewing these set points at any time with or without the output enabled

Active Down Programming

Standard internal circuitry to speed up response time to down programming changes

Universal AC input

Easy master/slave paralleling with active current sharing

Software

  • IVI-com driver for M130 option
  • Labwindows CVI driver for M9C option
  • Labview driver for M9C option

5 Year Warranty

規格說明
Common
Remote Sense The maximum load line drop is up to the full voltage rating of the supply. The drop in the load leads subtracts from the maximum voltage available for the load except as follows : maximum rated voltage is available at the load and voltage regulation specifications apply for line drops of < 1V on DLM 5-75 and DLM8-75, < 2V for all other DLM600 models
Remote Programming Voltage, current ( 0-100% ) and OVP ( 5-110% ) of full scale can be programmed by selectable 0-5VDC, 0-10VDC, or 0-5kΩ
Remote Monitoring Voltage or current can be monitored with user selectable ranges, scaled to 0 - 5 VDC or 0 - 10 VDC
Operational Features Master / slave parallel operation, up to 4 units of the same model can be connected in parallel, with active current sharing control. Series operation, multiple units of the same model can be connected in series, limited by 300Vpk between either output terminal and chassis.
Software IVI-COM, LabVIEW® or LabWindow® / CVI driver for Ethernet and IEEE-488.2
Regulatory Compliance CE Compliant :
• Low Voltage Directive ( 73 / 23 / EEC ) using EN 61010-1, and
• EMC Directive ( 89 / 336 / EEC ) using EN 61326 Certified to
UL 61010-1, CSA C22.2 No. 61010.1 and IEC/EN 61010-1

Input
Voltage and Frequency 90-132 VAC or 180-264 VAC max, auto ranging, 47-63 Hz, single phase, 2-wire plus ground
Current 11A maximum at 115VAC, 6A maximum at 230 VAC
Connector IEC 320 with detachable line cord
Power Factor 0.6, typical at full load; dependent on the impedance of the AC input.

Output
Line Regulation Voltage : 0.005% of V max + 2 mV
Current : 0.01% of I max + 2 mA
See table for input voltage variation over the AC input voltage range, with constant rated load.
Load Regulation Voltage : 0.005% of V max + 2 mV
Current : 0.02% of I max + 5 mA
See table For 0-100% load variation, with constant nominal line voltage
Transient Response 500 μs to steady-state output voltage ( within 0.1% of Vmax ) for 50-100% or 100-50% load change
Stability ±0.05% of maximum voltage or current over 8 hours after 30 minute warm-up time at fixed line, load and temperature
Efficiency 84% typical at maximum output power; 82% typical for DLM 5-75 and DLM 8-75
Temperature Coefficient 0.02% / °C of maximum output voltage, 0.03% / °C of maximum output current. Change in output per °C change in ambient temperature, with constant line and load.

Physical
Dimensions Width: 8.5” ( 216 mm )
Height: 1U or 1.75” ( 44 mm )
Depth: 17” ( 432 mm ) Option M6 depth 20” ( 508 mm )

Weight & Shipping Weight
Weight 9.7 lbs. ( 4.4 kg )
Shipping Weight 12.7 lbs. ( 5.8 kg )

Output Ratings
Model Voltage Current Power
DLM 5-75 0-5 VDC 0-75 ADC 375 W
DLM 8-75 0-8 VDC 0-75 ADC 600 W
DLM 10-60 0-10 VDC 0-60 ADC 600 W
DLM 20-30 0-20 VDC 0-30 ADC 600 W
DLM 40-15 0-40 VDC 0-15 ADC 600 W
DLM 60-10 0-60 VDC 0-10 ADC 600 W
DLM 80-7.5 0-80 VDC 0-7.5 ADC 600 W
DLM 150-4 0-150 VDC 0-4 ADC 600 W
DLM 300-2 0-300 VDC 0-2 ADC 600 W

Output Ratings
Model Line Regulation Load Regulation
Voltage (0.005% of Vmax + 2 mV) Current
(0.01% of Imax + 2 mA)
Voltage (0.005% of Vmax + 2 mV) Current
(0.02% of Imax + 5 mA)
DLM 5-75 2.25 mV 9.5 mA 2.25 mV 20 mA
DLM 8-75 2.4 mV 9.5 mA 2.4 mV 20 mA
DLM 10-60 2.5 mV 8 mA 2.5 mV 17 mA
DLM 20-30 3 mV 5 mA 3 mV 11 mA
DLM 40-15 4 mV 3.5 mA 4 mV 8 mA
DLM 60-10 5 mV 3 mA 5 mV 7 mA
DLM 80-7.5 6 mV 2.75 mA 6 mV 6.5 mA
DLM 150-4 9.5 mV 2.4 mA 9.5 mV 5.8 mA
DLM 300-2 17 mV 2.2 mA 17 mV 5.4 mA

Programming Accuracy 1
Ethernet ( M130/M131 )
Model Voltage (0.1% of VMax) Current ( 0.25% of IMax ) OVP ( 0.5% of 1.1 x VMax )
DLM 5-75 5 mV 188 mA 44 mV
DLM 8-75 8 mV 188 mA 44 mV
DLM 10-60 10 mV 150 mA 55 mV
DLM 20-30 20 mV 75 mA 110 mV
DLM 40-15 40 mV 38 mA 220 mV
DLM 60-10 60 mV 25 mA 330 mV
DLM 80-7.5 80 mV 19 mA 440 mV
DLM 150-4 150 mV 10 mA 825 mV
DLM 300-2 300 mV 5 mA 1.65 V

Programming Accuracy 1
GPIB
Model Voltage (0.1% of VMax) Current ( 0.25% of IMax ) OVP ( 0.5% of 1.1 x VMax )
DLM 5-75 16 mV 375 mA 55 mV
DLM 8-75 16 mV 375 mA 88 mV
DLM 10-60 20 mV 300 mA 110 mV
DLM 20-30 40 mV 150 mA 220 mV
DLM 40-15 80 mV 75 mA 440 mV
DLM 60-10 120 mV 50 mA 660 mV
DLM 80-7.5 160 mV 38 mA 880 mV
DLM 150-4 300 mV 20 mA 1.65 V
DLM 300-2 600 mV 10 mA 3.3 V

Display Accuracy
Model Voltage Current
DLM 5-75 35 mV 850 mA
DLM 8-75 50 mV 850 mA
DLM 10-60 60 mV 700 mA
DLM 20-30 200 mV 400 mA
DLM 40-15 300 mV 260 mA
DLM 60-10 400 mV 210 mA
DLM 80-7.5 500 mV 85 mA
DLM 150-4 850 mV 50 mA
DLM 300-2 2.5 V 30 mA

Ripple and Noise
Model Ripple and Noise, Voltage Mode OVP Adjustment
Range
(5% - 110% of
Vmax)
Ripple (RMS)* Noise (P-P)*
DLM 5-75 5 mV 30 mV 0.25 - 5.5 V
DLM 8-75 5 mV 30 mV 0.4 - 8.8 V
DLM 10-60 5 mV 30 mV 0.5 - 11 V
DLM 20-30 2.5 mV 15 mV 1 - 22 V
DLM 40-15 2.5 mV 15 mV 2 - 44 V
DLM 60-10 2.5 mV 20 mV 3 - 66 V
DLM 80-7.5 4 mV 20 mV 4 - 88 V
DLM 150-4 7 mV 40 mV 7.5 - 165 V
DLM 300-2 10 mV 60 mV 15 - 330 V

Ripple and Noise : Stability
Model Voltage ( 0.05% of VMax ) Current ( 0.05% of IMax )
DLM 5-75 2.5 mV 37.5 mA
DLM 8-75 4 mV 37.5 mA
DLM 10-60 5 mV 30 mA
DLM 20-30 10 mV 15 mA
DLM 40-15 20 mV 7.5 mA
DLM 60-10 30 mV 5 mA
DLM 80-7.5 40 mV 3.8 mA
DLM 150-4 75 mV 2 mA
DLM 300-2 150 mV 1 mA

Ripple and Noise
Model Traffic Coefficient Maximum Total Remote Sense Drop 
Voltage ( 0.02% / °C of Vmax ) Current ( 0.03% / °C of Imax )
DLM 5-75 1 mV / °C 22.5 mA / °C 1 V
DLM 8-75 1.6 mV / °C 22.5 mA / °C 1 V
DLM 10-60 2 mV / °C 18 mA / °C 2 V
DLM 20-30 4 mV / °C 9 mA / °C 2 V
DLM 40-15 8 mV / °C 4.5 mA / °C 2 V
DLM 60-10 12 mV / °C 3 mA / °C 2 V
DLM 80-7.5 16 mV / °C 2.25 mA / °C 2 V
DLM 150-4 30 mV / °C 1.2 mA / °C 2 V
DLM 300-2 60 mV / °C 0.6 mA / °C 2 V

Output Ratings
1 ANALOG-CONTROL Input 14 EXTERNAL-OFF Input (+)
2 EXTERNAL-OFF Return 15 Auxiliary 5 VDC Output (+)
3 OVP Programming Input 16 OVP Resistance Programming Output
4 REMOTE-CONTROL Status Output 17 OVP Resistance Programming Return
5 VOLTAGE-MODE Status Output 18 FAULT Status Output
6 Auxiliary 5 VDC Return 19 Voltage Monitor Output
7 Current Monitor Output 20 Voltage Resistance Programming Return
8 Voltage/Current Monitor Return 21 Voltage Resistance Programming Return
9 Voltage Programming Input 22 Voltage Resistance Programming Return
10 Current Programming Input 23 Voltage Resistance Programming Return
11 Not Used 24 LOCAL-LOCKOUT Input
12 V/I/OVP Programming Return 25 OVP Status Output
13 Not Used
1 - Readback accuracy is the same as programming accuracy for all parameters except GPIB voltage readback which is 0.25% of VMax
Loading...