產品介紹

Picotest Injector
  • Facebook
  • LINE
  • Twitter
  • LinkedIn
J2111B-Picotest 注入變壓器 DC-40MHz(PSRR) | 洛克Lockinc

J2111B

J2111B-Picotest Current Injector DC-40MHz(PSRR)

產品說明
  • High PSRR Low Noise Regulator with Universal input
  • Output Current Variation (Controllable) - 50mA
  • Total Output Current Variation (Controllable+Bias) - 74mA
  • 20nSec typ rise and fall time
  • DC-40MHz usable range (interconnection dependent)
  • Two Quadrant Bipolar operation works with positive or negative source
  • Built in offset for use with Network Analyzer
  • Precision current monitor with 50 Ohm output
  • Works with AWG, Function generator and network analyzer
  • Can be used to measure battery impedance
型錄下載 (332 KB)
手冊下載 (2.1 MB)
J2111B FAQs (371 KB)
功能說明

The current injector is one of the most versatile of the injector products. Coupled with the G5100A, or other equivalent function generator, it is capable of performing small signal load steps up to 40MHz, with very fast rising and falling edges. Using the G5100A, also allows the rise and fall times to be controlled, various waveforms or even arbitrary waveforms. This can be used to simulate the effects of many different types of loads, including high speed digital circuit loading, which is often largely dynamic.

The current injector can also be used to measure output impedance of power supplies, voltage regulators, power buses and even batteries. It can be used to non-invasively measure the stability of a combined input filter and the negative resistance of a switching power supply. It also has application in the measurement and extraction of transistor data, including small signal current gain, Ft and many other dynamic performance parameters.

 

In RF and instrumentation circuits it can be used to provide constant current biasing for class A amplifiers and buffers.

Measuring Output Impedance

The 3-port measurement is hard to calibrate given the nature of the setup and usage of the J2111A/B for current monitoring/sinking. The J2111A/B adds 25mA to load current which may be too large for some references.

It is generally a harder test to setup (often a current probe, which can be noisy is needed) and unlike the 1 or 2-port measurements you can't measure with power off (a useful data set in PDN design).

The current injector has two connections for the output, Output and GND. The input is DC+AC and can be connected to either a signal generator or a network analyzer. A built in bias current enables Class A operation for use with a network analyzer. The Current Injector and DC Bias injector can also be used for this purpose.

The output current is reduced 40dB from the input signal, resulting in 10mA/V scaling. The current monitor is designed to be terminated into 50 Ohms and can be used with the network analyzer, an oscilloscope or a DMM to monitor current.

When used in conjunction with a voltage probe, the analyzer can measure Voltage/Current, which is the impedance.

What's included:

1. J2111B Current Injector
2. J2170B High PSRR Power Adapter
3. User manual
4. 1 Year Warranty

規格說明
Specifications
Characteristic Typical Units
Output Current Variation (Controllable) 50 mA
Total Output Current Variation (Controllable+Bias) 74 mA
Max input voltage DC+AC +/-5 A / V
Absolute Output Voltage Requirements -60 to -1.5 and +1.5 to +60* V
Current Monitor 1 V / A
Modulation Input Impedance 50 ohms
Current Monitor Output Impedance 50 ohms
Modulator Gain 10m A / V
Offset Current (typical) +24m/0/-24m A
Usable Bandwidth DC-40M Hz
Temperature range 0-50 C
Maximum Altitude 6000 Ft
Absolute Maximum Voltage < 50VAC and 75VDC
 

* The J2111B may be connected to power supplies with output voltage ranging from -40V to -1V and 1V to 50V. Between -1V and +1V the J2111B will not work reliably. However, it is recommended that the output voltage remain above 1.5V for all specifications to be in compliance.J2111B increases the allowable voltage to -60 to -1.5 and +1.5 to +60 and improves thermal design.

 Caution: To avoid equipment damage and/or severe injuries or death ensure that the absolute maximum ratings are observed and not exceeded at all times.

Loading...